ExpressionMR400InstructionsforUseMonitoringNIBP1213
TochangetheNIBPalarmlimitsettings
Note
See chapter 4 for detailed alarm limit setting instructions and options.
5 Alarm limit, minimum
6 Lower alarm limit setting
7 Upper alarm limit setting
8 Alarm limit, maximum
9High button
10 Low button
11 NIBP Diastolic Alarm Limits menu label
12 NIBP Mean Alarm Limits menu label
Step Action
1 SelectthealarmlimitsettingsintheNIBPVSbox.
TheNIBPAlarmLimitsmenuappears.Currentsettingsare
displayed.
2 Selectthedesiredpressure:
NIBP(Sys)
NIBP(Dia)
NIBP(Mean)
Thepressure isselected.
3 SelecttheLowbuttonortheHighbutton.
Theselectedbutton
willbehighlightedandthecurrentadjustment
willbedisplayed.
4Usingthekeypad,ortheincrement,decrement,orOffbuttons,
enterthedesiredsetting.
Thecurrentadjustmentwillreflectthechange.
5PresstheEnterbuttontosavethesetting.
Thealarmlimitsettingisupdated.
6
Tochangetheremainingsettings,repeat
steps2,3,4and5.
Thecurrentadjustmentwillreflectthechange.