14‐8MaintenanceandTroubleshooting ExpressionMR400InstructionsforUse
Tocleanareusableaccessory
Step Action
1 Removetheaccessoryfromuse.
2 Removeallvisibledebrisfromtheaccessoryusingsoapandwater.
Never pour liquid onto the accessory.
3CleantheaccessorybythoroughlywipingitusingCaviWipes
disinfectanttowelettes.Discardtheusedtowelettes(refertoyour
facility'sbiohazardprocedurefordisposal).
Follow the Instructions for Use from the disinfectant manufacturer
to clean the probe.
4DisinfecttheaccessorybythoroughlywettingitusingCaviWipes
disinfectanttowelettes.Discardtheusedtowelettes(refertoyour
facility'sbiohazardprocedurefordisposal).
Follow the Instructions for Use from the disinfectant manufacturer
to disinfect the probe.
5Allowtheaccessorytodry.(Norinsingisrequired.)
6
Checkthe accessoryforanyresidualdebris.Ifanydebrisispresent,
repeatsteps2through5thenre‐examinetheitembefore
proceeding.
Disposable SPO2 attachments: After some use, adhesive residue may
accumulate at the fiber heads on the probe. Carefully remove any
residue with alcohol to keep the glass fiber ends clean.
7
Checkthe accessoryfordamage(cracks,holes,tears,cuts,et
cetera)anddiscardtheaccessoryifdamageisfound.
8Storetheaccessory;seepage2‐12fordetails.