A4Specifications ExpressionMR400InstructionsforUse
Battery
Type Cart:LithiumIon
Module:Lithiumpolymer
OperationTime
(Fullycharged)
Cart:Dependentuponenabledparametersandsettings:
•Alldisplays,alarms,andmonitoringfunctionscontinuously
for8hours
•ECG&SPO2continuouslyfor8hours
•CO2continuouslyfor6hours(withorwithoutAGENT)
•P1andP2continuouslyfor6hours
•AGENTanalysiscontinuously
for6hours
Temperaturecontinuouslyfor6hours
•NIBPreadingsevery5minutesfor6hours
Module:Approximately8hours
ChargeTime Cart:Torechargeafullydischargedbatteryisapproximately12
hours withtheMR400turnedoff.Batterychargetimeto90
percentofcapacityisapproximately6hours.
Module:Torechargea
fullydischargedbatteryisapproximately4
hours.Batterychargetimeto90percentofcapacityislessthan4
hours.
MinimumVoltage
(Fornormaloperation)
Cart:14.4V
Module:3.7V
Capacity Cart:75Wh
Module:3.1Wh
Environment
OperatingTemperatureRange 10to35°C(50to95°F)
StorageandTransportTemperatureRange Batteries:0to40°C(32to104°F)
Cart:‐20to60°C(4to140°F)
Wirelessmodulesandallotheraccessoriesnotspecifiedbelow:
20to60°C(4to140°F)
ECGskinprepgel:Followinstructionsontube
Quadtrodes:
10to32°C(50to90°F)
Transducerandcable(optional)(REF98 9803179721):‐15to60°C
(50to140°F)
O2sensor(AGENToption),StorageTemperature:+5to25°C(+41
to77°F);TransportTemperature‐40to50°C(40to122°F)
(Whenstoringortransportingintemperaturesbeyondthe
ranges
specifiedabove,removethedesignatedcomponentandstoreor
moveitappropriately.)
General

Table of Contents