A16Specifications ExpressionMR400InstructionsforUse
Accuracy*(includesstabilityanddrift) Halothane: ±0.15Vol.%at0to1.00Vol%
±0.20Vol%at1.00to5.00Vol%
Uns pec ified>5.00
Isoflurane: ±0.15Vol.%at0to1.00Vol%
±0.20Vol%at1.00to5.00Vol%
Uns pec ified>5.00
Sevoflurane: ±0.15Vol.%at
0to1.00Vol%
±0.20Vol%at1.00to5.00Vol%
±0.40Vol%at5.00to8.00Vol%
Uns pec ified>8.00
Desflurane: ±0.15Vol%at0to1.00Vol%
±0.20Vol%at1.00to5.00Vol%
±0.40Vol%at5.00to10.00Vol%
±0.60Vol%at10.00to15.00Vol%
±1.0Vol%at15.00
to18.00Vol%
Uns pec ified>18.00
Enflurane: ±0.15Vol.%at0to1.00Vol%
±0.20Vol%at1.00to5.00Vol%
Uns pec ified>5.00
CarbonDioxide: ±0.10Vol%at0to1.00Vol%
±0.20Vol%at1.00to5.00Vol%
±0.30Vol%at5.00to7.00Vol%
±0.50Vol%at7.00to10.00Vol%
Uns pec ified>10.00
NitrousOxide: ±2.00Vol%at0to20Vol%
±3.00Vol%at20.0to100Vol%
InterferenceGas CO
2:
N
2
O,O
2
,AnyAgent=0.1%
ABS
inaccuracyallowanceforeach
N
2
O:CO
2
,O
2
,AnyAgent=0.1%
ABS
inaccuracyallowanceforeach
Agents:CO
2
=0%
ABS
inaccuracyallowance
N
2
O,O
2
,2ndAgent=0.1%
ABS
inaccuracyallowanceforeach
FlowRate(fixed) 200±20ml/min(Adult,Pediatric)
150±15ml/min(Neonate)
Maximumspecifiedintervalforintervention
ofwater(hoursatspecifiedminimumsample
flowrate)
AGENTmode:Adultandpediatricis17hours@200ml/min,37°C,
100%RH;neonateis17hours@120ml/min,
37°C,10 0%RH
CO2mode:8hours@50mL/min+/‐10ml/min