CMU Base System CMU
1100.4903.12 6.8 E-10
STATus:PRESet Reset status registers
Command description FW vers.
This command sets the transition filters (PTRansition and NTRansition filters) and the ENABle regis-
ters of the STATus:OPERation and the STATus:QUEStionable registers to defined values:
PTRansition is set to 32767 (0x7FFF), i.e. all hardware events are detected and
transferred to the EVENt register.
NTRansition is set to 0, i.e. a hardware event that disappears does not cause any
change in the EVENt register.
The ENABle registers are also set to 0 so that events are not reported in the status byte.
V1.20