Optimizing the Image
5-32 LOGIQ 7 Online Help
Direction 2392536-100 Rev. 1
Max (Maximum) Enhance
Description Sets the acoustic output to its maximum setting (100%)
Values On/Off. When you deactivate Max Enhance, the acoustic output
is returned to its previous setting. Max Enhance is deactivated
by the system when you turn it off, change probes, or change
the contrast technique.
Benefits This control provides quick transition to High MI imaging. This
allows the user a quick one-button push to destroy the agent.
Useful when the user is interested in the bubble wash-in
characteristics of the anatomy being scanned.
Contrast Trigger
Description The Contrast Trigger scans images at set intervals, delaying
imaging according to the time delay that you specify.
Adjusting Press Trigger Off to activate the time delay. When you activate
the trigger, the Time Delay can be controlled via the dial beneath
the Touch Panel, Adjust the knob to set the time delay. The
Contrast Trigger is deactivated if you go into a Timeline Mode or
activate B Flow or Harmonics.
Values Values are 0.5, 1, 2, 3, 4, 5, 6, 7, 8, 9, and 10 seconds.
Sensitivity PRI
Description Sensitivity PRI is used to adjust the sensitivity of the Contrast
Agent. By increasing the sensitivity, you lower the frame rate; by
decreasing the sensitivity, you raise the frame rate.
Adjusting Adjust the dial left/right.
Values 1, 1.25, 1.5, 1.75 and 2.
Benefits By using Sensitivity PRI, sensitivity can be changed for each
anatomy of interest.