GYN Measurements
LOGIQ 7 Basic User Manual 9-91
Direction 2392206-100 Rev. 1
Follicle measurements
You can make left and right ovary follicle measurements from
one, two, or three distances.
One distance 1. To select the left or right, adjust the Side knob.
2. Select Follicle; an active caliper displays.
3. To position the active caliper at the start point, move the
Trackball.
4. To fix the start point, press Set.
The system fixes the first caliper and displays a second
active caliper.
5. To position the second active caliper at the end point, move
the Trackball.
A dotted line connects the measurement points.
6. To complete the measurement, press Set.
7. Press Clear.
The system displays the ovary follicle measurement in the
Results Window.