OB/GYN
9-92 LOGIQ 7 Basic User Manual
Direction 2392206-100 Rev. 1
Follicle measurements (continued)
Two distances 1. To select the left or right, adjust the Side knob.
2. Select Follicle; an active caliper displays.
3. Make the first distance measurement:
a. To position the active caliper at the start point, move the
Trackball.
b. To fix the start point, press Set.
The system fixes the first caliper and displays a second
active caliper.
c. To position the second active caliper at the end point,
move the Trackball.
A dotted line connects the measurement points.
d. To complete the measurement, press Set.
The system displays the distance value in the Results
Window and displays an active caliper.
4. To make the second distance measurement, repeat steps
a–d above.
5. Press Clear.
The system displays the ovary follicle measurement in the
Results Window.