GYN Measurements
LOGIQ 7 Basic User Manual 9-93
Direction 2392206-100 Rev. 1
Follicle measurements (continued)
Three distances 1. To select the left or right, adjust the Side knob.
2. Select Follicle; an active caliper displays.
3. Make the first distance measurement:
a. To position the active caliper at the start point, move the
Trackball.
b. To fix the start point, press Set.
The system fixes the first caliper and displays a second
active caliper.
c. To position the second active caliper at the end point,
move the Trackball.
A dotted line connects the measurement points.
d. To complete the measurement, press Set.
The system displays the distance value in the Results
Window. After the first and second distance
measurement, the system displays an active caliper.
4. To make the second and third distance measurement,
repeat steps a–d above.
After the third measurement, the system displays the ovary
follicle measurement in the Results Window.