Cardiology Measurements
LOGIQ 7 Basic User Manual 10-17
Direction 2392206-100 Rev. 1
Left Ventricle (continued)
One time interval
measurements
• Heart Rate, Teichholz
• Heart Rate for Two Chamber study
• Heart Rate for Four Chamber study
• Heart Rate for Two Chamber Area-Length study
• Heart Rate for Two Chamber Method of Disk study
• Heart Rate for Four Chamber Area-Length study
• Heart Rate for Four Chamber Method of Disk study
• Heart Rate for Bi-Plane Method of Disk study
Making a time
interval
measurement
1. Select the measurement.
The system displays an active caliper and a vertical dotted
line.
2. To position the caliper at the start point, move the Trackball.
3. To fix the first caliper, press Set.
The system displays a second active caliper.
4. To position the caliper at the end point, move the Trackball.
5. To complete the measurement, press Set.
The system displays the time interval in the Results
Window.