10-82 LOGIQ 7 Basic User Manual
Direction 2392206-100 Rev. 1
Stroke Volume Ratio (SV)
To measure stroke volume, you make a velocity measurement in
the Doppler spectrum. You measure an FCA (Flow Cross
Sectional Area) on the vessel in B-Mode. The system then uses
the velocity and FCA values to calculate SV (Stroke Volume).
1. Select SV from either the B-Mode or Doppler Touch Panel.
The trace caliper (horizontal dotted line) appears in the
Doppler spectrum.
2. Move the caliper to the measurement start point [dependent
on the trace method chosen (peak, floor, mean, or mode)].
3. Press Set.
The first caliper is fixed and a second is active.
4. Do one of the following:
If you are using auto trace, move the Trackball to
position the second caliper at the velocity measurement
end point. Press Set.
The waveform maximum velocities are automatically
traced and a velocity value is displayed.
If you are not using auto trace, move the Trackball to
trace the waveform. Press Set.
The trace waveform is fixed and the velocity value is
displayed.
A cross-point caliper appears in the B-Mode image to
measure FCA (Flow Cross Sectional Area).
5. Move the Trackball to position the caliper at a point on the
vessel wall.