Stress Echo
LOGIQ 7 Basic User Manual 10-91
Direction 2392206-100 Rev. 1
Starting acquisition
1. Select the template.
2. Press Begin/Cont.
3. Perform a scan that conforms with the view that is
highlighted in the template matrix on the Clipboard window.
4. Press the P1 (Image Store) key.
• If the actual stress level is configured to preview the
cine loop before storing, use the cine loop controls to
select the most appropriate heart cycle and if desired
adjust the loop markers. Press P1 again to save the
selected cine loop.
or
If you do not want to store the cine loop, press Freeze to
cancel. Return to the scan screen.
• If the actual stress level is not configured to preview the
cine loop before storing, the system automatically
stores the last heart cycle.
Stress levels can be configured for side by side display/
comparison of the reference loop from baseline or previous level
and the loop to acquire.
5. After storing the cine loop, the system automatically
highlights the next view in the matrix to be acquired.
6. Repeat previous steps until all required views are
completed.

Table of Contents

Other manuals for GE LOGIQ 7

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ 7 and is the answer not in the manual?

GE LOGIQ 7 Specifications

General IconGeneral
BrandGE
ModelLOGIQ 7
CategoryMedical Equipment
LanguageEnglish

Related product manuals