EasyManuals Logo

GE LOGIQ 7 Basic User Manual

GE LOGIQ 7
984 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #619 background imageLoading...
Page #619 background image
Stress Echo
LOGIQ 7 Basic User Manual 10-103
Direction 2392206-100 Rev. 1
Postponed image
assignment
The assignment of the cine loops to the view can be done on a
later stage on a stored Continuous Capture acquisition.
1. Perform the Continuous Capture. See ‘Activating
Continuous Capture’ for more information. (Steps 1 to 11).
2. Press Store all.
The entire Continuous Capture acquisition is stored. The
examination can be ended and the image assignment,
analysis and scoring can be done later.
3. Re-open the examination, if necessary.
4. Press Protocol. The Protocol screen displays.
5. Click the continuous capture images on the Protocol
Template screen.
The Continuous Capture selection screen displays.
6. Assign the cine loops to the view. See ‘Activating
Continuous Capture’ for more information. (Step 12 a - f).
7. Select Done.
8. Perform analysis and scoring.
Restart capture
from the
Continuous
Capture Selection
• Press Restart Capture.
The recording in memory is deleted and the Continuous
Capture starts again.
Resume
Continuous
Capture
• Press Continue Capture.
Resumes Continuous Capture recording (only if the
Continuous Capture buffer is not full).

Table of Contents