EasyManuals Logo

GE LOGIQ 7 Basic User Manual

GE LOGIQ 7
984 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #847 background imageLoading...
Page #847 background image
Test Patterns
LOGIQ 7 Basic User Manual 16-71
Direction 2392206-100 Rev. 1
Brightness Calibration (continued)
NOTE: After you calibrate the monitor you may need to adjust
peripheral settings.
Table 16-40: Typical Contrast Settings to Adjust the Monitor
Room Condition
Contrast for
PCVIC Parts No.
except 2349225
Contrast for
DGVIC PartsNo.
2349225
Contrast for
DGVIC PartsNo.
2349225-2
Dark Room for Radiology/Cardiology 50 45 50
Dim Room for Radiology/Cardiology 60 55 60
Bright Room for OB 70 65 70
Dark Room for Cardiology 60 55 60

Table of Contents

Other manuals for GE LOGIQ 7

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ 7 and is the answer not in the manual?

GE LOGIQ 7 Specifications

General IconGeneral
BrandGE
ModelLOGIQ 7
CategoryMedical Equipment
LanguageEnglish

Related product manuals