Probes and Biopsy
17-30 LOGIQ 7 Online Help
Direction 2392536-100 Rev. 1
Displaying the Guidezone (continued)
The needle may vary from the center line or guidezone for
various reasons:
• Needle barrel to needle clearance or strength.
• Bracket manufacturing tolerance.
• Needle deflection due to tissue resistance.
• Needle size chosen. Thinner needles may deflect more.
NOTE: Although the multi-angle guides are compatible with the Civco
Ultrapro and Ultrapro II, it is recommended the multi-angle
guides only be used with the Ultrapro II.
Table 17-14: LOGIQ 7 Biopsy Guide Availability
Fixed Angle
Multi-Angle
Probe MBX1 MBX2 MBX3
3C 4.0 6.0 8.0
3.5C/3.5CS 7.0
5C 7.0 4.0 5.5 7.0
M7C 3.5 5.5 7.5
E8C 5.0
7L 4.0 4.0 5.5 7.0
10L 2.0 2.0 4.0 6.0
M12L 1.5 2.5 3.5
3S 4.22 5.72 8.22
M3S 4.0 5.5 8.0
DANGER
Failure to match the guidezone displayed to the guide may
cause the needle to track a path outside the zone.
It is extremely important that when using the adjustable angle
biopsy guides, the angle displayed on the screen matches the
angle set on the guide, otherwise the needle will not follow the
displayed guidezone which could result in repeated biopsies or
patient injury.