This menu item is only available if under -[Operating parameter]- [Application]
"Standard" has been selected.
For the configuration the following order must be followed:
1. Defining an action
2. Determining a condition
3. Saving parameters
7.18.8.1 Defining an action
The possible actions are listed in the following table.
Selection list for actions of the inputs [Input 1/2/3 on/off]
Function SPM Bit Description
-no action- --- no function
set marker 1 X64 = 1 Set marker 1
set marker 2 X65 = 1 Set marker 2
set marker 3 X66 = 1 Set marker 3
select net X72 = 1 Select net
set zero X112 = 1 Set zero
set tare X113 = 1 Set tare
reset tare X114 = 1 Reset tare
set test X115 = 1 Activate the analog test
reset test X116 = 1 Finish the analog test
reset PWF X117 = 1 Reset power fail
set fixtare X118 = 1 Set fixtare (use the value in address D31
as a tare value)
get fixtare X119 = 1 Save gross value as fixtare in address
D31
clr marker 1 X64 = 0 Clear marker 1