12.2.5.10 Function number 24–29: Limit value (Read/Write)
Function number 24 Limit 1 on
Function number 25 Limit 1 off
Function number 26 Limit 2 off
Function number 27 Limit 2 off
Function number 28 Limit 3 on
Function number 29 Limit 3 off
12.2.5.11 Function number 30, 31: Fixed values (Read/Write)
Function number 30 Fixed value for analog output, value (num) 0…20000 corres-
ponds to 20 mA
Function number 31 Fixed value for preset tare, see also SetFixTare, GetFixTare
in Chapter
12.2.1.
12.2.5.12 Function number 80–92: state-controlled action bits (write)
Note:
For setting bits, see Chapter
12.2.2.3 .
Only setting and resetting of single bits is possible.
When changing a bit from 0 to 1, the corresponding action starts. After handling the
command, the bit must be reset. Application: The master writes cyclically.
The bit is set as Write_Value_Select with the specified number (see Chapter 12.2.2.3 ).
The bit is reset at the specified number +128.
Function number 80 SetZero Set the gross weight to zero.
Function number 81 SetTare The weighing point is tared.
Function number 82 ResetTare Reset tare.
Function number 83 SetTest Start the ADC test.
Function number 84 ResetTest Finish the ADC test.
Function number 85 ResetPwf Reset the PowerFail bit (function num-
ber 1; the bit was set after "power on").
Function number 86 SetFixTare Taring with weight in numerical address
D31 "FixTare".
Function number 87 GetFixTare The current gross weight is copied to
the numerical address D31.
Function number 89 ResetError The CmdError error bit is reset.
Function number 90 GetAlibi Read out selected record of the alibi
memory.
Function number 91 GetDate Read out current time and date.
X3 Process Indicator PR 5410 12 Fieldbus interface