CSIP Level 1 Troubleshooting: Known Symptoms, Causes, and Corrective Actions
Contained in this message are the software IDs of the incompatible boards. These IDs are inter-
nal software identifiers that are not related to any label on the PC board itself.
For TR boards, software ID “B1” is accepted by HD11 systems. If an EnVisor TR board ID (A1)
is reported in the error message, the system software has detected an incompatible TR board
installed in the system. This must be replaced with a correct board before the system can be
used.
For transducer connector modules, the following software IDs are accepted by HD11 systems:
Explora module ID = 3
Cartridge module ID = 2
Pencil probe module ID = 6