EasyManua.ls Logo

Philips HD11 - Figure 10-61 Removing the Power Supply

Philips HD11
413 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4535 612 10261 HD11 Service Manual Page 248
CSIP Level 1 Disassembly and Reassembly: Disassembly Procedures
Figure 10-61 Removing the Power Supply
Handle
10
Power supply ribbon cable
NOTE Ensure that the power supply
ribbon cable slides out with the
power supply. The cable shield-
ing may catch on the E-box
frame during this process.
10. Grasp the power supply handle and
firmly pull the power supply out of
the E-box housing.
This completes the power supply
removal procedure.
Perform this procedure in reverse for
reassembly.
Return to Disassembly Procedure List (2
of 2).

Table of Contents