4535 612 10261 HD11 Service Manual Page 248
CSIP Level 1 Disassembly and Reassembly: Disassembly Procedures
Figure 10-61 Removing the Power Supply
Handle
10
Power supply ribbon cable
NOTE Ensure that the power supply
ribbon cable slides out with the
power supply. The cable shield-
ing may catch on the E-box
frame during this process.
10. Grasp the power supply handle and
firmly pull the power supply out of
the E-box housing.
This completes the power supply
removal procedure.
Perform this procedure in reverse for
reassembly.
Return to Disassembly Procedure List (2
of 2).