4535 612 10261 HD11 Service Manual Page 211
CSIP Level 1 Disassembly and Reassembly: Disassembly Procedures
Figure 10-24 Removing the PC Access Panel
6
Handle
7
CAUTION Always observe proper
ESD procedures when
servicing the inside of the
PC. Wear a grounded
wrist strap and work on a
static-dissipating mat if
possible.
6. There are two twist-locks on the PC
access panel. Turn both the
twist-locks 90 degrees and pull the
access panel open using the handle
on the panel. Do not pull on the
twist-locks.
7. Slide the panel toward the front of
the system to fully remove it from
the PC assembly. Proceed to step 8.