4535 612 10261 HD11 Service Manual Page 78
CSIP Level 1 Installation: Inspecting the Installation Site
Start up and configure the system (page 83). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ❏
Set up the monitor (page 83). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ❏
Set the localization parameters (page 84 through page 93). . . . . . . . . . . . . . . . . . . . . . . . . ❏
Verify system options (page 94). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ❏
Configure the DICOM/connectivity network, if required (page 97).. . . . . . . . . . . . . . . . . . ❏
Configure other system settings (optional) (page 98). . . . . . . . . . . . . . . . . . . . . . . . . . . . . ❏
Prepare the peripherals (page 98). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ❏
Perform a system functional check (page 99). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ❏
Present the system (page 100). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ❏
Inspecting the
Installation
Site
Conduct an inspection of the site prior to system installation to ensure a trouble-free installation
and to assist the customer in providing an environment that supports reliable performance of the
system.
Checking
Physical Access
for Delivery and
Installation
Check that the installation site is physically acceptable as follows:
• With the customer, evaluate site access for delivery. Consider the availability of a loading
dock, elevators, ramps, and the width of passageways and doors.
• Verify that there is adequate space available for system uncrating and installation.
• At each location where the system will be operated, verify that there is enough space for the
system and any external peripherals.
• Check adjacent rooms for activity and types of equipment that might negatively affect the
ultrasound system.