4535 612 10261 HD11 Service Manual Page 43
CSIP Level 1 System Specifications: Introduction
2 System Specifications
Introduction
This section summarizes the system specifications for the HD11 Ultrasound System.
Physical
Specifications
Dimensions
Height (bottom of caster to top of monitor):
- Lowest position: <130.8 cm (51.5 in)
- Highest position: <148.6 cm (58.5 in)
Width: <55.8 cm (22 in)
Depth: <111.76 cm (44 in)
Weight
Less than 100 kg (220 lb) including monitor, and without peripherals.
Electrical
Specifications
AC input: 90 to 264 Vac, 47 to 63 Hz
Ground impedance: 100 mΩ maximum
Dielectric withstand:
- 1,500 Vac mains to safety ground
- 4,000 Vac mains to AC secondaries
Load: 1,150 VA maximum.
AC output (including PC): non-sinusoidal 120 VRMS, 350 VA maximum