4535 612 10261 HD11 Service Manual Page 138
CSIP Level 1 Performance Tests: Test and Inspection Matrices
Table 6-2 When to Perform Tests—Summary
Test Name
Test Code
(Ta bl e 6 -1 ) PM Tasks
Install/
Upgrade
Other
Service
Events Comments
Safety—Visual
Inspection
Sv:P Ye s Ye s S e e Ta b le 6 -3 . Not required on software-only
upgrades.
Safety—Ground
Continuity
Sgc(sn):P/x No Yes See Ta b le 6 -3 . Not required on software-only
upgrades.
Safety—System S:P/w/x/y/z Ye s No See Tab le 6 -3 . Not required on installations and
upgrades.
Safety—TEE and
Endocavity
Ste(sn):P/x Ye s No See Tab le 6 -3 .
Safety-Epiaortic/
Interoperative (CF)
Sep(sn):P/x/y Ye s No See Tab le 6 -3 .
Imaging Transducers Ti:P Yes Yes See Tab le 6 -3 .
Safety—
Transthoracic (BF)
Sbf(sn):P/x/y See
Comments
column
No See Ta b le 6 -3 . Test only if visual inspection
results in questionable electrical
integrity.
Non-Imaging
Transducers
Tn:P Yes Ye s S e e Tab l e 6 -3 .
Printers P:P Ye s Ye s S e e Ta b le 6 -3 .
VCRs V:P Ye s Ye s S e e Ta b le 6 -3 .
PC Xp:P Ye s Ye s S e e Ta b le 6 -3 .