4535 612 10261 HD11 Service Manual Page 129
CSIP Level 1 Performance Tests: Test and Inspection Matrices
Table 6-1 HD11 System Test and Inspection Matrix
Test Block
Name
(or Symbol)
Test or Inspection to Perform Expected Test Results
What to Record on
Service Record
Sv:
(System Visual
Inspection)
Inspect the system controls, cables, power
cord, transducer cables, lenses and controls,
and system and transducer's physical
condition. Are all of the above free from
physical damage that could make the
equipment unsafe?
Answer to question is “yes. Sv:P
or
Sv(sn):F
(sn = last 5 digits of
serial number,
P= Pass and F = Fail)
Sgc:
(System Ground
Continuity)
Measure the resistance between the ground
pin on the system power cord and the
system chassis ground.
< 200 milliOhms Sgc:P/x
or
Sgc(sn):F/x
(x = ohms measured,
and sn = last 5 digits of
serial number,
P= Pass and F = Fail)
S:
(System or
Stand-alone
Peripheral
Safety)
Measure with safety analyzer:
(w) Case (ground wire) leakage current
(normal condition)
(x) Case (chassis) ground resistance
w = < 300 a (UL—USA only),
or
< 500 μA (IEC/CSA)
x = < 200 mOhms (UL/IEC/
CSA)
S:P/w/x/y/z
or
S:F/w/x/y/z
(P = Pass, F = Fail, and
w, x, y, z = uA or
mOhms measured)