4535 612 10261 HD11 Service Manual Page 237
CSIP Level 1 Disassembly and Reassembly: Disassembly Procedures
Figure 10-50 Removing the Transducer Connector
Modules
Torx screws (4 plcs)
6
7
8
9
NOTE All of the transducer connec-
tor modules must be removed
before the Signal Distribution
Board can be removed.
6. Remove the four Torx screws secur-
ing the connector module bezel and
connector case.
7. Remove the module bezel.
8. Pull the connector case outward to
disconnect it from the Signal Distri-
bution Board and slide it out of the
E-box assembly.
9. Repeat step 6 through step 8 for the
remaining connector modules. Pro-