Philips HD11 - Figure 10-37 Removing the E-box Interface Board

Philips HD11
413 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4535 612 10261 HD11 Service Manual Page 224
CSIP Level 1 Disassembly and Reassembly: Disassembly Procedures
Figure 10-37 Removing the E-box Interface Board
3
E-box Interface
board
Hauppauge video
capture board
Physio Trigger cable
2
1
NOTE If you are removing the Haup-
pauge Video Capture board,
the E-box Interface board
should be removed first for
easier accessibility to the Video
Capture board.
To remove the E-box Interface
board
1. Disconnect the two cables con-
nected to the E-box Interface board
on the PC rear panel.
2. Disconnect the Physio Trigger cable.
3. Remove the one panhead screw
securing the board to the PC con-
nection slot.
This completes the E-box interface
board removal procedure.