4535 612 10261 HD11 Service Manual Page 224
CSIP Level 1 Disassembly and Reassembly: Disassembly Procedures
Figure 10-37 Removing the E-box Interface Board
3
E-box Interface
board
Hauppauge video
capture board
Physio Trigger cable
2
1
NOTE If you are removing the Haup-
pauge Video Capture board,
the E-box Interface board
should be removed first for
easier accessibility to the Video
Capture board.
➤ To remove the E-box Interface
board
1. Disconnect the two cables con-
nected to the E-box Interface board
on the PC rear panel.
2. Disconnect the Physio Trigger cable.
3. Remove the one panhead screw
securing the board to the PC con-
nection slot.
This completes the E-box interface
board removal procedure.