CSIP Level 1 Installation: Powering On and Configuring the System
Configuring the
DICOM Feature
If the system you are installing has the DICOM option, you or the customer will need to config-
ure the system DICOM feature. Since configuring the system for DICOM operation requires
information from the customer network administrator (see “Gathering Site Network (DICOM)
Information” on page 79), Philips recommends that you work with the customer to configure the
DICOM feature. The steps required to configure the system for DICOM communications over a
network are basically as follows:
To set or change DICOM parameters
1. Enter HD11 System DICOM Settings.
2. Enter Server Information and Assign Servers to Roles.
3. Set Up Automatic Export, Printing, Study Deletion, and Modality Worklist.
NOTE You must first set the system date, time, and time zone before configuring a DICOM
Modality Worklist Server. If you enable the Modality Worklist Server and then change
the system time, the patient list may not update properly.
For detailed information and instructions for configuring the DICOM feature, press ? Help on
the soft key panel after you have the system running and select “DICOM Setup” from the Help
contents. With the customer, follow the respective instructions in Help to configure the system
for DICOM operation.
This is a good time to verify that, in the DICOM Diagnostics window under System Log, the sys-
tem error logging level is set to Error Level Logging.