A-2 PN: 10580-00343 Rev. H S820E UG
A-1 Introduction Instrument Messages
Reset and then cycle the power to the instrument. If error message
persists, call your Anritsu Service Center.
9. Error executing remote query command: The remote query command
sent to the instrument has generated an unexpected error. Perform a
Factory Reset and then cycle the power to the instrument. If error
message persists, call your Anritsu Service Center.
10. Error occurred while zeroing. Please zero sensor without
RF input: When zeroing a USB power sensor, the RF input connector
of the sensor must not be connected to any source of RF power. If RF
power is detected during the zeroing, then the calibration will fail and
the zeroing will not be applied.
11. File cannot be recalled: File being recalled failed because the
instrument model used when the file was saved does not match the
current instrument model.
12. File missing: File being accessed cannot be found. The file has either
been deleted or moved to a different location.
13. Forward relative not applicable: For the USB In-line power sensor,
relative readings do not apply when making forward measurements of
CCDF.
14. Hardware driver failed to load: One of the hardware drivers required
has not loaded correctly. Perform a Factory Reset and then cycle the
power to the instrument. If error message persists, call your Anritsu
Service Center.
15. Loading USB Sensor ... : The USB power sensor has been detected and
a connection is being established to this sensor.
16. Model Mismatch: When recalling a setup file or measurement file, this
message is displayed if the file being recalled was generated by an
instrument that is not the same model as the one being used to recall
the file.
17. PLL Lock failed: Phase Lock Loop hardware has failed to lock. Perform
a Factory Reset and then cycle the power to the instrument. If error
message persists, call your Anritsu Service Center.
18. Recalling File...Failed: An error was encountered while recalling the
file. The recall process was aborted. Try again. If the error persists,
then the file may be corrupted.
19. Reverse relative not applicable: For the USB In-line power sensor,
relative readings do not apply when making reverse measurements of
Reflection Coefficient, Return Loss, or VSWR.