S820E UG PN: 10580-00343 Rev. H A-1
Appendix A — Instrument
Messages
A-1 Introduction
This appendix provides additional details regarding messages displayed on
the S820E Site Master. The text of the dialog boxes is shown below, listed in
alphabetical order.
Example Message:
Message Shown in Instrument: Additional details or suggestions
regarding the message.
1. Command to USB device returns error or invalid data: Check the
connection to the USB device and try again.
2. Device not ready: Device being accessed is not ready or responding. Try
to access again.
3. EEPROM corrupted: EEPROM device being accessed has been
corrupted. Perform a Factory Reset under the “Preset Menu”
on page 11-27 and then cycle the instrument power. If the error
message persists, call your Anritsu Service Center.
4. Encountered error in loading cable list. Default file restored:
The cable list file has been corrupted and will not be loaded. The
default file has been used.
5. Encountered error in loading keyboard settings. Default file
restored: The onscreen keyboard file has been corrupted and will not
be loaded. The default file has been used.
6. Encountered error in loading shortcut icon settings. Default file
restored: The menu shortcut file has been corrupted and will not be
loaded. The default file has been used.
7. Encountered error in restoring last setup. Default file restored: The
last setup file has been corrupted and will not be loaded. The default
file has been used.
8. Error executing remote command: The remote command sent to the
instrument has generated an unexpected error. Perform a Factory