EasyManua.ls Logo

Capintec CAPTUS 3000 - Chapter 10; Setup Counting Parameters

Capintec CAPTUS 3000
338 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
CAPINTEC, INC. CAPTUS
®
3000
Setup Counting Parameters
1. Click the
icon on Figure 10-2 Schilling Test Main Screen. The screen appears
as shown in Figure 10-3. This screen allows the user to enter counting parameters, which
will be the default settings each time a patient is entered into the Schilling Test patient
directory. These settings may be changed as needed for any individual patient or
patients.
Figure 10-3
Default Values frame
Standard Dilution Generally, the activity of the standard solution is too high to
measure in a well counter without dilution. A 1 to 100 dilution means that 1 part of
standard solution is mixed with 99 parts of non-radioactive solution to make a diluted
solution of 1%. A 1 to 50 dilution is a 2% solution, and so on. The default value is set
to 100.
Urine Sample Aliquot This is the volume used for the sample counted in the well
counter and should not exceed the useful counting dimensions of the well. An aliquot
sample volume of 4 ml is a good sample volume for the well detector provided with
the CAPTUS
®
3000 system.
Note: The aliquot volume must be the same for both the patient’s sample and the
standard solution sample.
Counting Time This is the duration for counting, in seconds, each time a
measurement is performed. The default value is set to 60 seconds.
Repetitions frame This is the number of repetitions each time a measurement is made.
For two repetitions, the average value is used in calculations. The second repetition may
be skipped during the measurement procedure, if so desired.
10-4 LAB TESTS November 13

Table of Contents

Related product manuals