CAPINTEC, INC. CAPTUS
®
3000
Measure Liquid and Multiply
by Factor
(Multiple Doses)
• The Lot Number, Calibration Date, Activity and Calibration Time may be
determined from your measurements of the dose in a dose calibrator or from the
previously prepared dose label.
• For Capsule Multiple Doses, enter the number of capsules to be administered
(Multiple Units Given). For Liquid Multiple Doses enter the Multiplying Factor [(Total
Vol of Liq) / (Vol of Liq Counted)].
Note: The fields in the Dosage Information section are available only when you
have selected a protocol for the patient.
4. Enter General Information.
• Click in the Probe Distance text box and input the distance of the Probe from the
patient to be measured in centimeters (cm) (10-50 cm).
• Click in the Counting Time text box and input the counting time in seconds (10-9,999
seconds).
• Click in the Tech ID text box and enter the alphanumeric ID (up to 32 characters) of
the person performing the procedure.
5. Click the
button to save the changes or click the button to cancel
the changes and return to Figure 7-1 Main Thyroid Uptake Screen.
6. To add more patients, click the
icon on Figure 7-1 Main Thyroid Uptake Screen.
• To move between data fields use the Tab key and to select from drop-down list
use ï‚ or  keys.
• To return to the CAPTUS
®
3000 Main screen, click the icon or use the
Viewing Patient Information
The Patient Status box displays a concise report of the highlighted patient record. If a test
has not yet been started, it will show the current data status.
To locate a specific patient record;
1. Click in the Patient Directory box to activate it.
7-12 THYROID UPTAKE November 13