CAPINTEC, INC. CAPTUS
®
3000
Note: If the system includes a Well, please verify that a Well Liner has been inserted into
the Well before use (refer to Figure 4-1 Well Counter Assembly). This prevents
contamination of the Well Detector if a spill or leak occurs.
Figure 4-1 Well Counter Assembly
ENVIRONMENT REQUIREMENTS
Operational
The instrument should be located where the level of the background radiation is as low and
as constant as possible.
The instrument should be located where the temperature is stable within a range of +50°F to
+85°F (+10°C to +30°C) and the maximum relative humidity is 90% non-condensing to
warrant maximum reliability and accuracy.
The instrument should be located where the barometric pressure is within a range of 27–31
inches of mercury (91–105 kilopascals).
Storage
The instrument should be stored where the temperature is stable and the range is from
+39°F to +110°F (+4°C to +43°C) and the maximum relative humidity is 90% non-condensing
to warrant maximum reliability.
The instrument should be stored where the barometric pressure is within a range of 15–33
inches of mercury (51–112 kilopascals).
CAUTION: If these environmental requirements are not followed, the instrument may
display erroneous readings.
4-2 GENERAL SETUP November 13

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Capintec CAPTUS 3000 and is the answer not in the manual?

Capintec CAPTUS 3000 Specifications

General IconGeneral
BrandCapintec
ModelCAPTUS 3000
CategoryMedical Equipment
LanguageEnglish

Related product manuals