CAPINTEC, INC. CAPTUS
®
3000
AUTO CALIBRATION
Calibration is done with Cs137 and Eu152 standard sources.
1. Click the button or press Alt+A on Figure 5-1 Quality Assurance Main
Screen. Figure 5-3 Auto Calibration Screen will appear. When the procedure is started,
the spectrum appears in the upper half of the screen as it is being collected and displays
the measurement time elapsed in the upper right corner. Note that the
button at the top of the screen is enabled only after calibration has been done at least
once on that day.
Figure 5-3 Auto Calibration Screen
2. Note that in the Select Detector frame, PROBE is selected by default. To change the
detector to Well, click the radio button marked WELL.
3. Click the button or press Alt+S to start the calibration procedure. The
screen appears as shown in Figure 5-4.
Figure 5-4
July 14 QUALITY ASSURANCE 5-5