CAPINTEC, INC. CAPTUS
®
3000
POWER REQUIREMENTS
CAUTION: If the input voltage is not within the stated limits, the unit may not
function correctly or may be damaged.
Input Power
Line Voltage: ............. 115 or 230 Vac depending on model ordered
Line Frequency: ........ 60 or 50 Hz
Current: ..................... 2.0 Ampere @ 115V
1.1 Ampere @ 230V
Fuse: For 115V use T 3.15AH, 250V, (Time Delay)
For 230V use T 2.5AH, 250V, (Time Delay)
Isolation Transformer Output
Computer and Monitor combined
Output Voltage: ......... 115Vac
Current Output: ......... 1.9A
Frequency: ................ 50/60Hz
Printer
Output Voltage: ......... 115Vac
Current Output: ......... 0.7A
Frequency: ................ 50/60Hz
TURN ON PROCEDURE
Verify all system components are mounted securely and all cables are connected. Connect
the Isolation Transformer main power cable to a medical grade wall socket.
All software necessary for proper operation of your CAPTUS
®
3000 Thyroid Uptake System
has been pre-installed and tested at the factory to ensure trouble-free setup.
Note: To remove main power from the system, the Isolation Transformer main power cable
must be unplugged from the wall socket.
There are three power switches associated with the system:
• the monitor has a power switch,
• the computer has a power switch, and
• the printer has a power switch.
Turn on all three switches.
Your system will power up, go through the computer boot-up sequence, and display the
Windows
®
start up screen.
November 13 GENERAL SETUP 4-3