ADDING A NEW ISOTOPE TO THE LIBRARY
Note: All fields in bold are required before the button will be enabled.
1. Click the
icon on Figure 6-1 Isotopes Main Screen. Figure 6-2 will appear.
Figure 6-2
2. Click in the appropriate text box under Isotope Information to input the isotope name
and its half-life. Select a time unit for the entered half-life from the drop-down list by
clicking the
arrow button in the Time Unit text box.
Note: An isotope name can be entered only in a specific format; which is the symbol of
the element (alphabetical character) followed by the atomic mass (number). If the
element symbol is one character, be sure to insert a space between the character
and the number. If the element symbol is two characters, do not insert a space
between the character and numbers. If entered otherwise, an error message will
appear as shown in Figure 6-3.
Figure 6-3
3. Click in the text boxes under Energy Information and input the peak energy values in
keV. At least one peak value must be input.
November 13 ISOTOPES 6-3

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Capintec CAPTUS 3000 and is the answer not in the manual?

Capintec CAPTUS 3000 Specifications

General IconGeneral
BrandCapintec
ModelCAPTUS 3000
CategoryMedical Equipment
LanguageEnglish

Related product manuals