To stop the counting midway and quit without saving, click the
button or
press Alt+C. The abort procedure screen appears as shown in Figure 10-19.
After count collection is completed, the screen appears as shown in Figure 10-22.
Figure 10-22
To accept the measurements and continue with the next step, click the
button or press the Y key. To discard and recount the current measurement, click the
button or press the N key.
Test Results
Once the measurement process for a patient is complete, clicking on the patient in the
Patient Directory on Figure 10-2 Schilling Test Main Screen will show the latest test result
in the Patient Status box. Test results are expressed as percent excretion, which is
calculated as follows:
( )
( )
DB-S
A
V
B-U100
Excretion %
×
××
=
where U = Urine sample count rate
B = Background (either room or patient blank)
S = standard count rate
V = total urine volume
A = urine aliquot
D = dilution of the standard
Patients Reports
1. Click on a patient in the Patient Directory box on Figure 10-2 Schilling Test Main
Screen to select a patient for viewing or printing the report.
2. Click the
icon. The screen appears as shown in Figure 10-23, showing a
typical patient report. It contains the patient’s demographic data, dosage and sample
information, count data and calculated results. If Normal Range values have been
November 13 LAB TESTS 10-15

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Capintec CAPTUS 3000 and is the answer not in the manual?

Capintec CAPTUS 3000 Specifications

General IconGeneral
BrandCapintec
ModelCAPTUS 3000
CategoryMedical Equipment
LanguageEnglish

Related product manuals