CAPINTEC, INC. CAPTUS
®
3000
Pre- Dose Measurement frame This option allows patient urine to be counted before
administering radioactive Vitamin B-12. The urine sample collected from the patient
before administration is called the patient blank. It will be subtracted from the post-dose
urine sample to yield net patient counts. This is an important measurement for a patient
who has previously received a radioactive isotope that may be excreted in the urine. It
may be skipped during the measurement procedure.
Normal Range of Excretion Frame Minimum and Maximum These numbers will
define a range of values for excretion that can be considered as medically normal. They
are entered as seen fit by the doctors. Out of range results indicate a medical condition
and will be flagged in the report. Entering these values is optional.
2. To input new values in any frame, click in the appropriate text box or radio button or use
the Tab key to move from one selection to another.
3. Click the
button to save changes or click the button to discard the
changes and return to Figure 10-2 Schilling Test Main Screen.
QUICK TIPS:
The Standard Dilution, Urine Sample Aliquot and Counting Time fields
appear again while adding a patient for a Schilling Test procedure. Any
changes made on the Add Patient screen will override the default values
entered in the Setup.
For greater flexibility, set Pre-Dose Measurement to Yes and Repetitions to
2, as these additional steps may be skipped if desired, while performing the
measurements.
To return to previous screen without saving changes press the ESC key.
Adding Patients
Note: All fields in bold are required before you can begin the Schilling Test. However, you
may input the Patient Name, Date of Birth and Patient ID alone, click the
button and input the remaining information later. The Patient Status box on Figure
10-2 Schilling Test Main Screen will display the message “Data Partially Entered”.
1. Click the
icon on Figure 10-2 Schilling Test Main Screen. The Patient
Information screen appears as shown in Figure 10-4, with information to be entered in
three sections.
November 13 LAB TESTS 10-5

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Capintec CAPTUS 3000 and is the answer not in the manual?

Capintec CAPTUS 3000 Specifications

General IconGeneral
BrandCapintec
ModelCAPTUS 3000
CategoryMedical Equipment
LanguageEnglish

Related product manuals