CAPINTEC, INC. CAPTUS
®
3000
ADDING WIPE LOCATIONS TO PRESET GROUPS
Note: All fields in bold are required before the button becomes enabled.
1. Click the
icon on Figure 8-2 Main Wipe Test Screen. Figure 8-5 Add Wipe
Location Screen will appear.
Figure 8-5 Add Wipe Location Screen
2. Input the wipe location name in the Location text box (e.g. Patient Waiting Area or Lab
Room No. 421, etc.). Any combination of up to 32 alphanumeric characters and spaces
may be used.
3. To select a wipe type (Preset Group), click the appropriate radio button in the Type
frame. Changing the Type will automatically apply Trigger Level and Counting Time
according to values entered in Figure 8-3 Wipe Test Setup Screen.
4. To change the Trigger Level or Counting Time for this wipe location, click in the
appropriate text box and type in the new number.
5. To add comments about the wipe location, click the
button or
press Alt+L.
Note: The comments entered will be displayed in the Wipe Comment box immediately
below the Wipe Locations box on Figure 8-2 Main Wipe Test Screen when the
location is selected. The comments will also appear in the Wipe Test results.
6. To add more wipe locations, click the
button or press Alt+N.
8-6 WIPE TESTS November 13

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Capintec CAPTUS 3000 and is the answer not in the manual?

Capintec CAPTUS 3000 Specifications

General IconGeneral
BrandCapintec
ModelCAPTUS 3000
CategoryMedical Equipment
LanguageEnglish

Related product manuals