CAPINTEC, INC. CAPTUS
®
3000
PRINTING MCA DATA
To print an MCA spectrum report, select the button on Figure 13-1 MCA Main
Screen. The printout will contain the spectrum shown on the screen, with ID information, date
and time of acquisition, real and live times, gain, voltage, and detector settings. For each
active ROI, total counts and total count rate, as well as starting and ending channels, integral
counts, and peak centroid channel will be printed.
CUSTOMIZING MCA SETTINGS
The MCA module provides full control over MCA settings used to acquire a spectrum. The
initial settings, determined in Autocalibration, are automatically transferred to the MCA
module, so that the 1024 channels are calibrated at about 2 keV/channel. However, these
can be changed using the following procedures.
Note: When the MCA mode is exited, the automatic calibration settings are restored. This is
to assure correct definition of any ROIs used in the clinical modules of the CAPTUS
®
3000.
To alter the automatic calibration settings, select the
button from the menu
bar located at the top of Figure 13-1 MCA Main Screen. The Calibration Settings frame will
appear in the MCA Main Screen as shown in Figure 13-16.
Figure 13-16
13-14 MCA November 13

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Capintec CAPTUS 3000 and is the answer not in the manual?

Capintec CAPTUS 3000 Specifications

General IconGeneral
BrandCapintec
ModelCAPTUS 3000
CategoryMedical Equipment
LanguageEnglish

Related product manuals