Quality Assurance Tests
Note: All QA Tests (Enter Check Source and Chi-Square Setup, Auto Calibration, Measure
Efficiency, Constancy Test, and Chi-Square Test) must be performed before any
measurements can take place. Measurement Setups (Thyroid Uptake Protocols,
Wipe Test Defaults, Wipe Test Groups, Bioassay) need only be performed if required.
Before performing any of the Quality Assurance tests, allow the system to warm up at least
30 minutes.
1. Click the
button beside “Enter Check Source and Chi-Square Setup”.
2. Enter the required fields. For details on the screen, see the pages listed in Reference
Pages below.
3. Upon exiting the Check Source and Chi-Square Setup screen, you will be automatically
returned to the Startup Tips screen.
4. Continue with the next test in the list (e.g., Auto Calibration).
5. Repeat steps 1-4 for each remaining test (Measure Efficiency, Constancy Test, and Chi-
Square Test).
4-14 GENERAL SETUP November 13