CAPINTEC, INC. CAPTUS
®
3000
Figure 4-11 Report Header Frame
6. Click the
button or press Alt+D to save the changes or click the
button or press Alt+C to exit without saving the changes. Figure 4-9 Startup Tips
Window will re-appear.
QUICK TIPS:
To move between boxes use the Tab key and to move within boxes use or
keys.
To customize your report printouts, you can input the address of your medical
facility as the report header.
Quality Assurance Tests
Note: All QA Tests (Enter Check Source and Chi-Square Setup, Auto Calibration, Measure
Efficiency, Constancy Test, and Chi-Square Test) must be performed before any
measurements can take place. Measurement Setups (Thyroid Uptake Protocols,
Wipe Test Defaults, Wipe Test Groups, Bioassay) need only be performed if required.
Before performing any of the Quality Assurance tests, allow the system to warm up at least
30 minutes.
1. Click the
button beside “Enter Check Source and Chi-Square Setup”.
2. Enter the required fields. For details on the screen, see the pages listed in Reference
Pages below.
3. Upon exiting the Check Source and Chi-Square Setup screen, you will be automatically
returned to the Startup Tips screen.
4. Continue with the next test in the list (e.g., Auto Calibration).
5. Repeat steps 1-4 for each remaining test (Measure Efficiency, Constancy Test, and Chi-
Square Test).
4-14 GENERAL SETUP November 13

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Capintec CAPTUS 3000 and is the answer not in the manual?

Capintec CAPTUS 3000 Specifications

General IconGeneral
BrandCapintec
ModelCAPTUS 3000
CategoryMedical Equipment
LanguageEnglish

Related product manuals