Located by each line cord on the Isolation Transformer are labels (Figure 1-6) indicating
which device each cord is to be connected to and the power availability for that device.
POWER
CORD
COMPUTER
POWER
PRINTER
CORD
CORD
POWER
MONITOR
50/60 Hz
115V @ 0.7A
50/60 Hz
115V @ 1.9A
COMBINED
Figure 1-6
CAUTIONS AND NOTES
CAUTION: Changing the date or time while a timed test is in progress (for example,
a 24 hour Thyroid Uptake Test) might affect the results or introduce
unexpected error. Care should be taken to evaluate any potential impact
for ongoing tests before changing the date or time. Verify that in the Date
and Time Properties in Windows Control Panel that the “Automatically
adjust clock for daylight saving changes” checkbox is not checked.
CAUTION: Only qualified/trained personnel should operate or service this unit.
CAUTION: If the equipment is used in a manner not specified in this manual, the
protection provided by the equipment may be impaired.
CAUTION: High voltage is present inside the Detectors (up to 1000 Volts). Due to
the presence of this high voltage, opening the covers with the system
plugged in may be hazardous. Refer all servicing to qualified personnel.
CAUTION: There are no internal adjustments inside the detector, computer or
printer that should be performed by the user within the conditions of the
warranty, except for changing the Isolation Transformer fuse. Due to the
presence of high voltages, opening the cover of the computer with the
system plugged in may be hazardous. Refer all servicing to qualified
personnel.
CAUTION: Never use the Well Detector without the well liner in place. Liners are
inexpensive and easy to replace. A contaminated detector is a very
costly mistake. If unit becomes contaminated remove liner and clean unit
as stated in CHAPTER 15: CLEANING AND MAINTENANCE before
operating.
November 13 SAFETY 1-5