CAPINTEC, INC. CAPTUS
®
3000
CAUTION: To remove main power from the system, the Isolation Transformer power
cable must be unplugged from the medical grade wall socket.
Note: It is recommended that periodic (every five years) re-calibration of the unit be
performed only by Capintec’s Authorized Service Center (reference CHAPTER 15:
CLEANING AND MAINTENANCE) to guarantee that the instrument's high reliability is
maintained).
GENERAL SAFETY TIPS
• Unplug the equipment before cleaning it. Use only a damp cloth; do not use solvents
or aerosol cleaners.
• To protect the equipment from overheating, make sure none of the openings (vents)
on the computer, monitor or the printer are blocked.
• Do not use the equipment directly in front of a radiator, heat register or air
conditioning duct.
• Do not use the equipment near water or spill liquids of any kind into the equipment.
• Be sure that your power source matches the rating listed on the Isolation
Transformer.
• The CAPTUS
®
3000
power cord has a grounded, 3-prong plug as a safety feature,
and it will only fit into a grounded outlet. Do not use an adapter to defeat the
grounding.
• The CAPTUS
®
3000 computer system is connected to the Isolation Transformer to
comply with electrical safety for Medical Electrical Equipment.
• Due to the risk of electrical shock, do not install or use a modem or a network adapter
with the CAPTUS
®
3000 computer system.
• To avoid damaging the power cord, do not place anything on it or place it where it will
be walked on. If the cord becomes damaged, replace it immediately.
• Aside from the routine maintenance described in this manual, do not try to service
this equipment yourself. There are no user serviceable parts contained in the system
other than the fuses in the Isolation Transformer. Do not make any adjustments other
than those outlined in this manual, as you may in-validate the calibration or cause
damage requiring extensive repair work. Refer servicing to qualified service
personnel.
• Do not subject the detectors to rough handling, or rapid temperature changes, as the
NaI Scintillation Crystal can be damaged.
November 13 SAFETY 1-7

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Capintec CAPTUS 3000 and is the answer not in the manual?

Capintec CAPTUS 3000 Specifications

General IconGeneral
BrandCapintec
ModelCAPTUS 3000
CategoryMedical Equipment
LanguageEnglish

Related product manuals