GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
Chapter 2 - Site Preparations 2 - 13
2-3-5 Networking  Setup Requirements
2-3-5-1 Stand Alone Scanner (without Network Connection)
None.
2-3-5-2 Scanner Connected to Hospital’s Network
Supported networks:
DICOM Network (option)
2-3-5-3 Purpose of the DICOM Network Function
DICOM services provide the operator with clinically useful features for moving images and patient 
information over a hospital network. Examples of DICOM services include the transfer of images to 
workstations for viewing or transferring images to remote printers. As an added benefit, transferring 
images in this manner frees up the on-board monitor and peripherals, enabling viewing to be done while 
scanning continues. With DICOM, images can be archived, stored, and retrieved faster, easier, and at 
a lower cost.
2-3-5-4 DICOM Option  Setup Requirements
To configure the Vivid 7 to work with other network connections, the site’s network administrator must 
provide information to complete the form in Figure 2-2 "Pre-quote Worksheet for DICOM Network 
Information" on page 2-14. Ensure that there are no spaces in any field of the form. 
Entries must include: 
• A host name, local port number, AE Title, IP address and Net Mask for the Vivid 7. 
• The IP addresses for the default gateway and other routers at the site for ROUTING 
INFORMATION.
• The host name, IP address, port and AE Title for each device the site wants connected to the Vivid 7
 
for DICOM APPLICATION INFORMATION. A field for the make (manufacturer) and the revision of 
the device, is also included. This information may be useful for error solving.