DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL
Chapter 7 - Diagnostics/Troubleshooting 7 - 63
Section 7-8
Acquisition Diagnostics
7-8-1 Acquisition Diagnostics Introduction
If the scanner is running, but an error that may be a hardware failure, has been registered, then
continue to read.
7-8-1-1 Start System Test
To start the System Test, power up the scanner, see 4-2-2 - Power ON/ Boot UP on page 4-3 for detailed
instructions.
After scanner initialization is complete, continue with the steps below;
1.) Press CONFIG
(F2) and log on as adm, see 4-2-4 "Log On to the System as ‘ADM’" on page 4-12.
2.) Select SYSTEM
from the bottom of the menu that appears on the monitor.
3.) Select TEST
.
.
4.) Select GO TO SYSTEM TEST to select the Calibration/Test screen.
Figure 7-67 System > Test selection screen