GE HEALTHCARE
DIRECTION FC091194, REVISION 11    VIVID 7 SERVICE MANUAL 
5 - 98 Section 5-4 - Back-End Processor (BEP)
5-4-10 UPS Battery Description
The UPS Battery is used in these Back-End Processors:
• BEP1
• BEP2.0
• BEP2.2
• BEP4 used as BEP2.2 Replacement
• BEP4 used as BEP2.0 and BEP1 Replacement
The UPS batteries provide power to the Back-End Processor to enable a controlled shut down of the 
processor. 
NOTE: It is not possible to use the Vivid 7 for scanning or any other useful work without the mains 
power. The UPS battery is only used as a power source for the BEP so it can do a controlled 
shut down, or be in standby modus without AC Mains Power connected for a limited period of 
time, maximum 15 minutes.
Figure 5-58   Battery Block Diagram
Figure 5-59   Battery Pack
In Line
20A Fuse
Battery 1
Battery 2
Compression tape used to
minimize battery pack movement 
while installed in chassis
Heat-shrink holds batteries together
and insulates terminals.
Spacer
seperates
batteries
+
-
-
+
Label to be
attached to
this surface
In line 20A fuse