GE HEALTHCARE
DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL
3 - 92 Section 3-10 - Connectivity Setup - Software v3.x
3-10-6 Set the Remote Archive’s Network Information - Software Version v3.x
In the Remote Archive Setup area of the screen (see example in Figure 3-95), enter the;
1.) Remote Archive IP address. (Default IP Address from factory: 10.0.0.4).
2.) Remote Archive Name. (Default Remote Archive Name from factory: ECHOPAC7-000001).
.
3-10-7 Save the New Settings
1.) Press SAVE SETTINGS to save the new settings.
.
2.) The new settings are saved to a common settings file. After a restart, the settings are also included
in other screens.
3.) Restart Vivid 7 to activate the changes.
Figure 3-95 TCP/IP Set-up for Vivid 7
Figure 3-96 Save New TCP/IP settings
IP Address for the remote archive
Computer Name for the remote archive
Save settings