GE HEALTHCARE
DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL
Chapter 8 - Replacement Procedures 8 - 91
Section 8-14
Rear Casters Replacement Procedure
8-14-1 Manpower
2 persons, 30 minutes (one Caster)
8-14-2 Tools
Phillips screwdriver size 2
5 mm hexagon key
Bevel Edged Board and Wooden Wedge (service parts, see Table 9-77 "Service Kits, Overview" on
page 9-99.)
8-14-3 Preparations
1.) Shut down the system, see 4-2-3 "Power Shut Down" on page 4-7.
2.) Remove the Side Covers, see "Side Covers Removal Procedure" on page 8-4.
3.) Remove Filter Cover, see "Upper Rear Cover Removal Procedure" on page 8-6.
4.) Remove Lower Rear Cover, see "Lower Rear Cover Removal Procedure" on page 8-10.
Figure 8-86 Bevel Edged Board and Wooden Wedge.
Wooden Wedge
Bevel Edged Board