10 - 16 Section 10-6 - Using a Phantom
10-7-5 Chassis Leakage Current Test
10-7-5-1 Definition
This test measures the current that would flow in a grounded person who touched accessible metal 
parts of the bedside station if the ground wire should break. The test verifies the isolation of the power 
line from the chassis. The meter is connected from accessible metal parts of the case to ground. 
Measurements should be made with the unit On and Off, with the power line polarity Normal and 
Reversed. Record the highest reading. 
10-7-5-2 Generic Procedure
The test verifies the isolation of the power line from the chassis. The testing meter is connected from 
accessible metal parts of the case to ground. Measurements should be made with the unit ON and OFF, 
with the power line polarity Normal and Reversed. Record the highest reading of current.
When using the Microguard or a similar test instrument, its power plug may be inserted into the wall 
outlet and the equipment under test is plugged into the receptacle on the panel of the meter. This places 
the meter in the grounding conductor and the current flowing from the case to ground will be indicated 
in any of the current ranges. The maximum allowable limit for chassis source leakage is shown in Table 
10-12 "Chassis Leakage Current Limits—Accessible Metal Surfaces" on page 10-13.
7.) Follow the test conditions described for respective test points shown in Table 10-15 "Chassis 
Leakage Current Test Condition" on page 10-17.
CAUTION
Electric Shock Hazard. When the meter's ground switch is OPEN, don't touch the unit!
CAUTION
Equipment damage possibility. Never switch the Polarity and the status of Neutral when the unit 
is powered ON. Be sure to turn the unit power OFF before switching them using the POLARITY 
switch and/or the NEUTRAL switch. Otherwise, the unit may be damaged.
Figure 10-3   Set Up for Chassis Source Leakage Current, IEC 601-1 Clause 19 - Continuos Leakage Currents 
and Patient, Auxiliary Currents