GE HEALTHCARE
DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL
Chapter 3 - System Setup 3 - 93
3-10-8 Connect to a DICOM Server in a Network - Software Version v3.x
In this case the Vivid 7 is configured to work with DICOM servers in a network environment. Images are
first saved on the local image buffer on the scanner. At the end of the examination the images are sent
to the DICOM server via a DICOM spooler and to the local database, depending on dataflow.
This scenario requires that the scanner is configured to be connected to DICOM servers as described
below.
3-10-8-1 Overview
To work against the DICOM server, the following information has to be entered in the scanner:
• The DICOM server IP address.
• The DICOM server port number.
• The DICOM server AE title (the server application’s name).
• Name of Device.
3-10-8-2 DICOM Server IP Address Setting on the Scanner - Software Version v3.x
1.) Press CONFIG
(F2) and log on as adm, see 4-2-4 "Log On to the System as ‘ADM’" on page 4-12.
2.) Select CONNECTIVITY
(in the lower part of the window).
3.) Select the DATAFLOW
tab.
4.) Select the arrow to the right for the Name field to list all dataflows in a pull-down menu, see Figure
3-97.
5.) Select the dataflow you want to configure from the Dataflow Name pull-down list, see Figure 3-97.
Figure 3-97 Select Dataflow
Arrow
Select dataflow from
Dataflow Name pull-down
menu.