GE HEALTHCARE
DIRECTION FC091194, REVISION 11 VIVID 7 SERVICE MANUAL
Chapter 8 - Replacement Procedures 8 - 61
8-9-9 Backup of Patient Database & User Presets - Software Version v2.x (cont’d)
5.) Select Backup now. A Warning will appear on the screen, as shown in Figure 8-55 on page 8-61.
.
6.) Select OK.
NOTE: No process indicator will appear on screen. Verify that the “Finished, OK” Status appears in the
Result column, and that the date appearing in the Last Successful Backup column is the current
date and time. The backup is completed.
7.) Remove the Backup MO Disk when the backup is completed.
Figure 8-55 Backup Warning